    MROZ & Heatsan system for Northeast hospital acute dialysis clinic

    A large hospital in the Northeast had an antiquated central water system for their acute dialysis clinic, this system required significant maintenance and oversight. The hospital sought to upgrade the central water system and to optimize the manpower associated with the operation and maintenance of the system.

    There were some constraints with regards to upgrading the hospital’s system, this included:

    • It was not going to be possible to shut down the central system to install the new water system due to the clinic’s operating requirements.
    • The water room space was limited, as the room size was less than 100 ft2.

    AmeriWater proposed the installation of a new central system featuring an open frame MROZ Reverse Osmosis system and a Heatsan system for heat disinfection of the water loop.

    The design of the AmeriWater system offered the following features and benefits to address the hospital’s requirements:

    • The modular design of the MROZ & Heatsan components allowed the system to fit into the limited space of the water room area.
    • The use of the Heatsan system reduced the required manpower associated with loop disinfection. The Heatsan system operates automatically through a programmable controller and for this system it was programmed to disinfect the loop on Sunday mornings, starting at 1 AM.

    To address the need for operating the central system while the new system was being installed, the AmeriWater dealer ran a temporary DI loop. The running of the temporary DI loop, removal of the old water system, and installation/start-up of the new water system was accomplished within a two-week period.

    The system installed by AmeriWater meets all AAMI water specifications, and since start-up has been operating reliably and without maintenance issues.