AmeriWater Capsule Sterile Processing Water System – 800 – 2400 GPD
Water quality in ambulatory surgery centers (ASC) plays a critical role. High standards of cleanliness must be maintained to avoid potential fines. While this is motivation to follow a higher standard of care, the safety and well-being of patients, should always be the primary driver.
By implementing AmeriWater’s ASC all-in-one water system, facilities will experience optimal ST108 water quality and prolonged lifespans for medical equipment.
This reliable water purification solution is designed with a simplified, space saving footprint and ensures patient well-being and essential ANSI/AAMI ST108:2023 standards are met.
- Minimum 25% Space Savings
- Plug and Play Water In-and-Out
- Regulatory Assurance
- Simplified Installation and Maintenance
- Lower Startup and Plumbing Costs
Features
Applications
Specifications
Certifications
Accessories
Downloads
Features
- Anti-Scalant Injection Feed to Prevent Hard Water Scale on Reverse Osmosis Membranes
- AWRO System with:
– Smart monitoring with real-time alarms and system shutdowns
– Programable standby flush
– High-efficiency membranes for top-tier water quality
– Open frame design for ease of serviceability - Polish DI with Silex Deionizer
– Easy DIY resin exchange
– Light indicator signals time for resin exchange - Bypass Options for Continuous Operation During Maintenance or System Shutdown
- Storage Tank with Sealed Lid
– Distribution pump sized to meet system requirements with shut down on empty storage tank
– Submicron vent filter
– Distribution loop return manifold
– Pressure relief valve
– Distribution loop return flow meter
– Sample port
– Distribution loop to drain valve
– Internal spray nozzle - Ultraviolet Light
– Controls bacteria proliferation in purified water storage and distribution systems
– Equipped with an online monitor to measure output intensity and alarms when lamp needs replaced - Endotoxin Ultrafiltration
– 0.2 Submicron Endotoxin Cartridge
– Installed post UV to remove endotoxins from the purified water storage and distribution system - Real-Time Remote Monitoring Enabled
– Sends mobile and email alerts
– Key system tracking
– Online dashboard that stores system data and runs reports - Included Installation Kit, Water System Identification Kit, and AAMI ST108 Validation Water Testing
- System Can Be Easily Disinfected
Applications
Additional Capsule Views
Front Right Angle
- Storage Tank
- Ozone Port
- Flow Meter
- Remote Monitoring Panel
- RO
- Sediment Filter
- DI
- Anit-Scalant

Side View
- UV
- Endotoxin
- Membrane Housing

Back View
- Endotoxin
- Power Box
- Loop Out
- Loop Return
- Feedwater
- Anti-Scalant Pump

Specifications
- Drain: Floor Drain or Sink Must be Within 4′ of Unit and Able to Accommodate a Flow Rate of 10 GPM
- Feed Water Quality: Potable Drinking Water Standards
- Feed Water Pressure Min./Max.: 20/80 PSI Dynamic
- Projected Rejection: 95% – 98%
- Recovery: 50%
- Silica: 40 ppm
- Silt Density Index: < 5
- Turbidity: < 1 NTU
- pH Range: 6.0 – 8.5
- Iron: 0.1 ppm
| Models | 00SPDVER800 | 00SPDVER2400 |
| RO Production Rated Capacity* | 800 GPD | 2,400 GPD |
| Number of Membranes | 2 | 1 |
| Storage Tank Volume | 55 Gallons | 55 Gallons |
| Distribution Pump Output | 10 GPM @ 30 PSI | 30 GPM @ 30 PSI |
| Minimum Feed Flow | 2 GPM | 6 GPM |
| Feedwater Temperature | 77°F (Blend Valve Recommended) | 50°F – 90°F (77°F Nominal Temperature) |
| Dimensions in Inches (WxDxH) | 32 x 41.5 x 67 | 33.5 x 51 x 79 |
| * Product flow rate varies with temperature and inlet pressure. All models are rated at 77°F with feed water of 1500 mg/l NaCl @ 55 PSI feed pressure and pH of 7.5. As a safety factor use 60% of the rated capacity when sizing or add blend valve to temper the feed water. | ||
Certifications
- ETL Listed, Conforms to UL Std 61010-1
- Certified CAN/CSA C22.2 No. 61010-1
- ISO 13485 Registered Device
Accessories
Downloads
Your All-in-One SPD and Surgery Center Water System
Download >